Abstract

Purpose: Neuromelanin MRI (NM-MRI) is applied as a proxy measurement of dopaminergic functioning of the substantia nigra pars compacta (SN). To increase its clinical applicability, a fast and easily applicable NM-MRI sequence is needed. This study therefore compared accelerated NM-MRI sequences using standard available MRI options with a validated 2D gradient recalled echo NM-MRI sequence with off-resonance magnetization transfer (MT) pulse (2D-MToffRes). Methods: We used different combinations of compressed sense (CS) acceleration, repetition times (TR), and MT pulse to accelerate the validated 2D-MToffRes. In addition, we compared a recently introduced 3D sequence with the 2D-MToffRes. Results: Our results show that the 2D sequences perform best with good to excellent reliability. Only excellent intraclass correlation coefficients were found for the CS factor 2 sequences. Conclusion: We conclude that there are several reliable approaches to accelerate NM-MRI, in particular by using CS.
Original languageEnglish
JournalNeuroradiology
Early online date2022
DOIs
Publication statusE-pub ahead of print - 2022

Keywords

  • Acceleration
  • Neuromelanin
  • Reliability
  • Substantia nigra

Cite this