TY - JOUR
T1 - Evaluation of delivered dose for a clinical daily adaptive plan selection strategy for bladder cancer radiotherapy
AU - Lutkenhaus, Lotte J.
AU - Visser, Jorrit
AU - de Jong, Rianne
AU - Hulshof, Maarten C. C. M.
AU - Bel, Arjan
PY - 2015
Y1 - 2015
N2 - To account for variable bladder size during bladder cancer radiotherapy, a daily plan selection strategy was implemented. The aim of this study was to calculate the actually delivered dose using an adaptive strategy, compared to a non-adaptive approach. Ten patients were treated to the bladder and lymph nodes with an adaptive full bladder strategy. Interpolated delineations of bladder and tumor on a full and empty bladder CT scan resulted in five PTVs for which VMAT plans were created. Daily cone beam CT (CBCT) scans were used for plan selection. Bowel, rectum and target volumes were delineated on these CBCTs, and delivered dose for these was calculated using both the adaptive plan, and a non-adaptive plan. Target coverage for lymph nodes improved using an adaptive strategy. The full bladder strategy spared the healthy part of the bladder from a high dose. Average bowel cavity V30Gy and V40Gy significantly reduced with 60 and 69ml, respectively (p <0.01). Other parameters for bowel and rectum remained unchanged. Daily plan selection compared to a non-adaptive strategy yielded similar bladder coverage and improved coverage for lymph nodes, with a significant reduction in bowel cavity V30Gy and V40Gy only, while other sparing was limited
AB - To account for variable bladder size during bladder cancer radiotherapy, a daily plan selection strategy was implemented. The aim of this study was to calculate the actually delivered dose using an adaptive strategy, compared to a non-adaptive approach. Ten patients were treated to the bladder and lymph nodes with an adaptive full bladder strategy. Interpolated delineations of bladder and tumor on a full and empty bladder CT scan resulted in five PTVs for which VMAT plans were created. Daily cone beam CT (CBCT) scans were used for plan selection. Bowel, rectum and target volumes were delineated on these CBCTs, and delivered dose for these was calculated using both the adaptive plan, and a non-adaptive plan. Target coverage for lymph nodes improved using an adaptive strategy. The full bladder strategy spared the healthy part of the bladder from a high dose. Average bowel cavity V30Gy and V40Gy significantly reduced with 60 and 69ml, respectively (p <0.01). Other parameters for bowel and rectum remained unchanged. Daily plan selection compared to a non-adaptive strategy yielded similar bladder coverage and improved coverage for lymph nodes, with a significant reduction in bowel cavity V30Gy and V40Gy only, while other sparing was limited
U2 - https://doi.org/10.1016/j.radonc.2015.06.003
DO - https://doi.org/10.1016/j.radonc.2015.06.003
M3 - Article
C2 - 26096847
SN - 0167-8140
VL - 116
SP - 51
EP - 56
JO - Radiotherapy and oncology
JF - Radiotherapy and oncology
IS - 1
ER -