TY - JOUR
T1 - EXTENT study
T2 - Medium-term outcomes of EXTra-design engineering inner-branch ENdografts for the treatment of complex aortic aneurysms from a multicenter collaboration
AU - Abisi, Said
AU - Zayed, Hany
AU - Frigatti, Paolo
AU - Furlan, Federico
AU - Simonte, Gioele
AU - Isernia, Giacomo
AU - Kuczmik, Wacław
AU - Fattoum, Maher
AU - Halak, Moshe
AU - Silverberg, Daniel
AU - Gkoutzios, Panos
AU - Saha, Prakash
AU - EXTENT Collaborators
AU - Waldeck, Stephan
AU - Becker, Benjamin V.
AU - Buz, Semih
AU - Sharma, Paritosh
AU - Matson, Matthew
AU - Hamady, Mohamed
AU - Riga, Celia
AU - Hoksbergen, Arjan
AU - Yeung, Kak Khee
AU - Lovato, Luigi
AU - Buia, Francesco
AU - Marzo, Ana Cristina
N1 - Publisher Copyright: © 2024 Society for Vascular Surgery
PY - 2024
Y1 - 2024
N2 - Objective: This study aims to present the medium-term outcomes of Extra-Design engineering endografts with inner branches (EDE-iBEVARs, Artivion) in endovascular aortic repairs of complex aneurysms building upon promising early results. Methods: A retrospective, international, multi-center study was conducted including consecutive patients who underwent complex endovascular aortic repairs using EDE-iBEVARs between 2018 and 2022. Patient demographics, aneurysm anatomical features, procedural details, reinterventions, complications, and endograft failures during follow-up were assessed. The primary outcome was aneurysm-related mortality. Secondary outcome measures included the freedom from all-cause mortality and reintervention, technical and clinical success, and late related complications including branch instability, endoleaks, and serious adverse events. Results: Our study encompassed a total of 260 patients across 13 European centers. The cohort included patients with thoracoabdominal aortic aneurysms (n = 116), suprarenal or juxta-renal aneurysms (n = 95), and those who had previous open repair or previous endovascular aortic repair with type 1A endoleak (n = 49). Of 982 possible inner branches (937 antegrade and 45 retrograde), 962 (98%) were successfully cannulated and bridged with covered stents during the index procedure. Overall, the endograft was successfully deployed in 98% of patients, and 93% were discharged from hospital following surgery. At 3 years, freedom from aneurysm-related mortality was 97%, whereas the freedom of all-cause mortality was 89%. Freedom from reinterventions was 91% and 76% at 1 and 3 years, respectively. The rate of late complications such as endoleaks or branch instability events was 12% (n = 30). The late branch occlusion rate during follow-up was 1.5% (n = 15), of which 12 were renal branches. Conclusions: EDE-iBEVARs demonstrate satisfactory medium-term outcomes with reintervention rates comparable to other endografts. Encouragingly, rates of branch patency were high, and major adverse events were low. This technology could expand the treatment options for patients with challenging complex aortic conditions.
AB - Objective: This study aims to present the medium-term outcomes of Extra-Design engineering endografts with inner branches (EDE-iBEVARs, Artivion) in endovascular aortic repairs of complex aneurysms building upon promising early results. Methods: A retrospective, international, multi-center study was conducted including consecutive patients who underwent complex endovascular aortic repairs using EDE-iBEVARs between 2018 and 2022. Patient demographics, aneurysm anatomical features, procedural details, reinterventions, complications, and endograft failures during follow-up were assessed. The primary outcome was aneurysm-related mortality. Secondary outcome measures included the freedom from all-cause mortality and reintervention, technical and clinical success, and late related complications including branch instability, endoleaks, and serious adverse events. Results: Our study encompassed a total of 260 patients across 13 European centers. The cohort included patients with thoracoabdominal aortic aneurysms (n = 116), suprarenal or juxta-renal aneurysms (n = 95), and those who had previous open repair or previous endovascular aortic repair with type 1A endoleak (n = 49). Of 982 possible inner branches (937 antegrade and 45 retrograde), 962 (98%) were successfully cannulated and bridged with covered stents during the index procedure. Overall, the endograft was successfully deployed in 98% of patients, and 93% were discharged from hospital following surgery. At 3 years, freedom from aneurysm-related mortality was 97%, whereas the freedom of all-cause mortality was 89%. Freedom from reinterventions was 91% and 76% at 1 and 3 years, respectively. The rate of late complications such as endoleaks or branch instability events was 12% (n = 30). The late branch occlusion rate during follow-up was 1.5% (n = 15), of which 12 were renal branches. Conclusions: EDE-iBEVARs demonstrate satisfactory medium-term outcomes with reintervention rates comparable to other endografts. Encouragingly, rates of branch patency were high, and major adverse events were low. This technology could expand the treatment options for patients with challenging complex aortic conditions.
KW - Endovascular aortic repair
KW - Inner branch endografts
KW - Thoracoabdominal aneurysms