T1 - Inhalation anesthesia and shielding devices to allow accurate preclinical irradiation of mice with clinical linac-based systems: Design and dosimetric characteristics
T2 - Design and dosimetric characteristics
AU - Lagerweij, Tonny
AU - Sewing, Charlotte
AU - van Battum, Leo
AU - Koken, Phil
AU - Heukelom, Stan
N1 - Publisher Copyright: © 2020 The Author(s) Copyright: Copyright 2021 Elsevier B.V., All rights reserved.
PY - 2021/1/1
Y1 - 2021/1/1
N2 - This technical note describes two devices to enable accurate irradiation of mice on clinical linac-based systems. To study the effects of radiation in murine, preclinical animal models, controlled and accurate dosing is important. This is not only important when specific volumes need to be irradiated, but also when the whole animal body is irradiated. To enable both purposes, we designed two devices. One device to administer Total Body Irradiation (TBI) simultaneously to six, free walking mice, and a second device, denoted as target box, in which we irradiate specific parts of the mice whilst organs-at-risk (OAR) are protected. In this latter device, we can position the mice in multiple ways. One configuration allows to sedate twelve mice simultaneously by isoflurane inhalation anesthesia and protect the body by lead shielding to allow radiation of the head only. Alternatively, the target box can be used to sedate maximal 4 mice simultaneously to irradiate the flank or paws only. All these setups allow high experimental throughput and thus a minimal occupation of the clinical equipment. As measured, the delivered radiation dosages in the regions of interest were accurate for both devices. In this technical note, we describe the design and build of these devices.
AB - This technical note describes two devices to enable accurate irradiation of mice on clinical linac-based systems. To study the effects of radiation in murine, preclinical animal models, controlled and accurate dosing is important. This is not only important when specific volumes need to be irradiated, but also when the whole animal body is irradiated. To enable both purposes, we designed two devices. One device to administer Total Body Irradiation (TBI) simultaneously to six, free walking mice, and a second device, denoted as target box, in which we irradiate specific parts of the mice whilst organs-at-risk (OAR) are protected. In this latter device, we can position the mice in multiple ways. One configuration allows to sedate twelve mice simultaneously by isoflurane inhalation anesthesia and protect the body by lead shielding to allow radiation of the head only. Alternatively, the target box can be used to sedate maximal 4 mice simultaneously to irradiate the flank or paws only. All these setups allow high experimental throughput and thus a minimal occupation of the clinical equipment. As measured, the delivered radiation dosages in the regions of interest were accurate for both devices. In this technical note, we describe the design and build of these devices.
KW - Animal models
KW - Fractionated radiation
KW - Inhalation anesthesia
KW - Linac
KW - Small animal irradiation devices
UR - http://www.scopus.com/inward/record.url?scp=85097771482&partnerID=8YFLogxK
U2 - https://doi.org/10.1016/j.ctro.2020.11.012
DO - https://doi.org/10.1016/j.ctro.2020.11.012
M3 - Article
C2 - 33367118
SN - 2405-6308
VL - 26
SP - 92
EP - 97
JO - Clinical and Translational Radiation Oncology
JF - Clinical and Translational Radiation Oncology
ER -