Long-Term Clinical Outcomes of Subcutaneous Versus Transvenous Implantable Defibrillator Therapy

Tom F. Brouwer, Dilek Yilmaz, Robert Lindeboom, Maurits S. Buiten, Louise R. A. Olde Nordkamp, Martin J. Schalij, Arthur A. Wilde, Lieselot van Erven, Reinoud E. Knops

Research output: Contribution to journalArticleAcademicpeer-review

137 Citations (Scopus)

Abstract

BACKGROUND: Transvenous implantable cardioverter-defibrillators (TV-ICDs) improve survival in patients at risk for sudden cardiac death, but complications remain an important drawback. The subcutaneous ICD (S-ICD) was developed to overcome lead-related complications. Comparison of clinical outcomes of both device types in previous studies was hampered by dissimilar patient characteristics.

OBJECTIVES: This retrospective study compares long-term clinical outcomes of S-ICD and TV-ICD therapy in a propensity-matched cohort.

METHODS: The authors analyzed 1,160 patients who underwent S-ICD or TV-ICD implantation in 2 high-volume hospitals in the Netherlands. Propensity matching for 16 baseline characteristics, including diagnosis, yielded 140 matched pairs. Clinical outcomes were device-related complications requiring surgical intervention, appropriate and inappropriate ICD therapy, and were reported as 5-year Kaplan-Meier rate estimates.

RESULTS: All 16 baseline characteristics were balanced in the matched cohort of 140 patients with S-ICDs and 140 patients with TV-ICDs (median age 41 years [interquartile range: 30 to 52 years] and 40% women). The complication rate was 13.7% in the S-ICD group versus 18.0% in the TV-ICD group (p = 0.80). The infection rate was 4.1% versus 3.6% in the TV-ICD groups (p = 0.36). Lead complications were lower in the S-ICD arm compared with the TV-ICD arm, 0.8% versus 11.5%, respectively (p = 0.03). S-ICD patients had more nonlead-related complications than TV-ICD patients, 9.9% versus 2.2%, respectively (p = 0.047). Appropriate ICD intervention (antitachycardia pacing and shocks) occurred more often in the TV-ICD group (hazard ratio [HR]: 2.42; p = 0.01). The incidence of appropriate (TV-ICD HR: 1.46; p = 0.36) and inappropriate shocks (TV-ICD HR: 0.85; p = 0.64) was similar.

CONCLUSIONS: The complication rate in patients implanted with an S-ICD or TV-ICD was similar, but their nature differed. The S-ICD reduced lead-related complications significantly, at the cost of nonlead-related complications. Rates of appropriate and inappropriate shocks were similar between the 2 groups.

Original languageEnglish
Pages (from-to)2047-2055
Number of pages9
JournalJournal of the American College of Cardiology
Volume68
Issue number19
DOIs
Publication statusPublished - 8 Nov 2016

Keywords

  • Adult
  • Aged
  • Arrhythmias, Cardiac/therapy
  • Death, Sudden, Cardiac/epidemiology
  • Defibrillators, Implantable
  • Female
  • Follow-Up Studies
  • Humans
  • Incidence
  • Male
  • Middle Aged
  • Netherlands/epidemiology
  • Retrospective Studies
  • Survival Rate/trends
  • Treatment Outcome