Optical coherence tomography of the Ex-press™ miniature glaucoma implant

Research output: Chapter in Book/Report/Conference proceedingConference contributionAcademicpeer-review

Abstract

Description of the use of Optical Coherence Tomography to localize the Ex-PRESS™ mini implant placed in the anterior segment of the eye. An OCT scanner, central wavelength 1280 nm, bandwidth 60 nm, resolution of 12 μm, was build onto a slitlamp to scan the anterior segment of the eye. Five ex-vivo porcine eyes received an Ex-Press implant, and were used as a model to visualize the position of the implant in the anterior segment. In the ex-vivo porcine eyes the OCT images showed the anatomy of the anterior segment in great detail. The anterior segment OCT was able to visualize the whole outline, and the position of the implant. The acquisition time of 0.8 sec is short enough to allow for scanning patients, and AS-OCT is expected to aid in providing answers to the question which parameters will determine the success or failure of such a device. © 2005 Optical Society of America.
Original languageEnglish
Title of host publicationEuropean Conference on Biomedical Optics, ECBO 2005
PublisherOptical Society of America (OSA)
ISBN (Print)0819458635
DOIs
Publication statusPublished - 2005
EventEuropean Conference on Biomedical Optics, ECBO 2005 - , Germany