TY - JOUR
T1 - Quantitative Method for the Analysis of Ivacaftor, Hydroxymethyl Ivacaftor, Ivacaftor Carboxylate, Lumacaftor, and Tezacaftor in Plasma and Sputum Using Liquid Chromatography With Tandem Mass Spectrometry and Its Clinical Applicability
AU - Vonk, Steffie E. M.
AU - van der Meer-Vos, Marloes
AU - Amsterdam Mucociliary Clearance Disease (AMCD) research group
AU - Bos, Lieuwe D. J.
AU - Neerincx, Anne H.
AU - Majoor, Christof J.
AU - Maitland-van der Zee, Anke-Hilse
AU - Mathôt, Ron A. A.
AU - Kemper, E. Marleen
N1 - Publisher Copyright: Copyright © 2021 The Author(s). Published by Wolters Kluwer Health, Inc. on behalf of the International Association of Therapeutic Drug Monitoring and Clinical Toxicology. Copyright: This record is sourced from MEDLINE/PubMed, a database of the U.S. National Library of Medicine
PY - 2021/8/1
Y1 - 2021/8/1
N2 - BACKGROUND: The novel cystic fibrosis transmembrane conductance regulator (CFTR) modulators, ivacaftor, lumacaftor, and tezacaftor, are the first drugs directly targeting the underlying pathophysiological mechanism in cystic fibrosis (CF); however, independent studies describing their pharmacokinetics are lacking. The aim of this study was to develop a quantification method for ivacaftor and its 2 main metabolites, lumacaftor and tezacaftor, in plasma and sputum using liquid chromatography with tandem mass spectrometry. METHODS: The developed method used a small sample volume (20 µL) and simple pretreatment method; protein precipitation solution and internal standard were added in one step to each sample. Liquid chromatography with tandem mass spectrometry was performed for a total run time of 6 minutes. The method was validated by assessing selectivity, carryover, linearity, accuracy and precision, dilution, matrix effects, and stability. RESULTS: The selectivity was good as no interference from matrices was observed. In the concentration range from 0.01 to 10.0 mg/L, calibration curves were linear with a correlation coefficient >0.9997 for all compounds. The within-run and between-run accuracy were between 99.7% and 116% at the lower limit of quantitation (LLOQ) and between 95.8% and 112.9% for all concentrations above LLOQ for all analytes in plasma and sputum. Within-run and between-run precisions were <12.7% for LLOQ and <6.7% for the higher limit of quantitation. Samples were stable, with no significant degradation at examined temperatures and time points. Clinical applicability was revealed by analyzing samples from 2 patients with CF. CONCLUSIONS: The presented method enables simultaneous quantification of ivacaftor, lumacaftor, and tezacaftor in plasma and sputum and is an improvement over previous methods because it uses smaller sample volumes, a simple pretreatment protocol, and includes tezacaftor. In future studies, it can be applied for examining pharmacokinetics characteristics of new CF transmembrane conductance regulator modulators.
AB - BACKGROUND: The novel cystic fibrosis transmembrane conductance regulator (CFTR) modulators, ivacaftor, lumacaftor, and tezacaftor, are the first drugs directly targeting the underlying pathophysiological mechanism in cystic fibrosis (CF); however, independent studies describing their pharmacokinetics are lacking. The aim of this study was to develop a quantification method for ivacaftor and its 2 main metabolites, lumacaftor and tezacaftor, in plasma and sputum using liquid chromatography with tandem mass spectrometry. METHODS: The developed method used a small sample volume (20 µL) and simple pretreatment method; protein precipitation solution and internal standard were added in one step to each sample. Liquid chromatography with tandem mass spectrometry was performed for a total run time of 6 minutes. The method was validated by assessing selectivity, carryover, linearity, accuracy and precision, dilution, matrix effects, and stability. RESULTS: The selectivity was good as no interference from matrices was observed. In the concentration range from 0.01 to 10.0 mg/L, calibration curves were linear with a correlation coefficient >0.9997 for all compounds. The within-run and between-run accuracy were between 99.7% and 116% at the lower limit of quantitation (LLOQ) and between 95.8% and 112.9% for all concentrations above LLOQ for all analytes in plasma and sputum. Within-run and between-run precisions were <12.7% for LLOQ and <6.7% for the higher limit of quantitation. Samples were stable, with no significant degradation at examined temperatures and time points. Clinical applicability was revealed by analyzing samples from 2 patients with CF. CONCLUSIONS: The presented method enables simultaneous quantification of ivacaftor, lumacaftor, and tezacaftor in plasma and sputum and is an improvement over previous methods because it uses smaller sample volumes, a simple pretreatment protocol, and includes tezacaftor. In future studies, it can be applied for examining pharmacokinetics characteristics of new CF transmembrane conductance regulator modulators.
UR - http://www.scopus.com/inward/record.url?scp=85112123529&partnerID=8YFLogxK
U2 - https://doi.org/10.1097/FTD.0000000000000829
DO - https://doi.org/10.1097/FTD.0000000000000829
M3 - Article
C2 - 33165217
SN - 0163-4356
VL - 43
SP - 555
EP - 563
JO - Therapeutic drug monitoring
JF - Therapeutic drug monitoring
IS - 4
ER -