T1 - The assessment of environmental and external cross-contamination in preparing ready-to-administer cytotoxic drugs
T2 - a comparison between a robotic system and conventional manual production
AU - Werumeus Buning, Annabel
AU - Geersing, Tjerk H.
AU - Crul, Mirjam
N1 - Funding Information: This research had not been funded by any specific grant from any funding agency in the public, commercial or not‐for‐profit sectors. Publisher Copyright: © 2019 The Authors. International Journal of Pharmacy Practice published by John Wiley & Sons Ltd on behalf of Royal Pharmaceutical Society Copyright: Copyright 2020 Elsevier B.V., All rights reserved.
PY - 2020/2/1
Y1 - 2020/2/1
N2 - Objectives: The primary aim of the study was to compare environmental and external (cross-) contamination of traces of cytostatics, during preparation of 5-fluorouracil and cyclophosphamide using a robotic system (APOTECAchemo) or the conventional manual compounding procedure. The secondary aim was to validate the cleaning procedure of the robot. Methods: Eighty ready-to-administer (RTA) infusion bags with 5-fluorouracil, cyclophosphamide or sodium chloride were compounded using both techniques on 3–5 days. Wipe samples were taken from several locations in the compounding room before and after cleaning, and also from the technician’s gloves. These samples were analysed for 5-fluorouracil and cyclophosphamide concentrations using GC/MS/MS. Key findings: A total of 284 wipe samples were collected during the study (113 from the manual and 171 from the robotic process). External contamination on the outside of infusion bags was 3.75% for both manual and robotic compounding. For manual compounding, external cross-contamination occurred on 2.5% of the prepared infusion bags. External cross-contamination occurred on 1.25% of the infusion bags for the robotic procedure. Inside the compounding room, 9% of the environmental wipe samples were contaminated in case of manual production and 24% for robotic compounding. Since 50% of the contaminated environmental samples for the robotic system were taken after cleaning, the cleaning procedure was extended and parameter setting for cyclophosphamide handling was performed. After this, residual environmental or external contamination was no longer detectable. Conclusion: Comparison of both preparation methods showed that external (cross-)contamination of infusion bags was lower using the robotic system. An optimized cleaning procedure showed the best results in environmental contamination for the robot.
AB - Objectives: The primary aim of the study was to compare environmental and external (cross-) contamination of traces of cytostatics, during preparation of 5-fluorouracil and cyclophosphamide using a robotic system (APOTECAchemo) or the conventional manual compounding procedure. The secondary aim was to validate the cleaning procedure of the robot. Methods: Eighty ready-to-administer (RTA) infusion bags with 5-fluorouracil, cyclophosphamide or sodium chloride were compounded using both techniques on 3–5 days. Wipe samples were taken from several locations in the compounding room before and after cleaning, and also from the technician’s gloves. These samples were analysed for 5-fluorouracil and cyclophosphamide concentrations using GC/MS/MS. Key findings: A total of 284 wipe samples were collected during the study (113 from the manual and 171 from the robotic process). External contamination on the outside of infusion bags was 3.75% for both manual and robotic compounding. For manual compounding, external cross-contamination occurred on 2.5% of the prepared infusion bags. External cross-contamination occurred on 1.25% of the infusion bags for the robotic procedure. Inside the compounding room, 9% of the environmental wipe samples were contaminated in case of manual production and 24% for robotic compounding. Since 50% of the contaminated environmental samples for the robotic system were taken after cleaning, the cleaning procedure was extended and parameter setting for cyclophosphamide handling was performed. After this, residual environmental or external contamination was no longer detectable. Conclusion: Comparison of both preparation methods showed that external (cross-)contamination of infusion bags was lower using the robotic system. An optimized cleaning procedure showed the best results in environmental contamination for the robot.
KW - automated compounding
KW - cytotoxic drugs
KW - environmental contamination
KW - hazardous drugs
KW - robotic preparation
KW - robotic system
KW - surface contamination
KW - wipe sampling
UR - http://www.scopus.com/inward/record.url?scp=85071867167&partnerID=8YFLogxK
U2 - https://doi.org/10.1111/ijpp.12575
DO - https://doi.org/10.1111/ijpp.12575
M3 - Article
C2 - 31489970
SN - 0961-7671
VL - 28
SP - 66
EP - 74
JO - International Journal of Pharmacy Practice
JF - International Journal of Pharmacy Practice
IS - 1
ER -