Correction for the Hematocrit Bias in Dried Blood Spot Analysis Using a Nondestructive, Single-Wavelength Reflectance-Based Hematocrit Prediction Method

Sara Capiau, Leah S. Wilk, Pieter M. M. de Kesel, Maurice C. G. Aalders, Christophe P. Stove

Research output: Contribution to journalArticleAcademicpeer-review

49 Citations (Scopus)

Abstract

The hematocrit (Hct) effect is one of the most important hurdles currently preventing more widespread implementation of quantitative dried blood spot (DBS) analysis in a routine context. Indeed, the Hct may affect both the accuracy of DBS methods as well as the interpretation of DBS-based results. We previously developed a method to determine the Hct of a DBS based on its hemoglobin content using noncontact diffuse reflectance spectroscopy. Despite the ease with which the analysis can be performed (i.e., mere scanning of the DBS) and the good results that were obtained, the method did require a complicated algorithm to derive the total hemoglobin content from the DBS's reflectance spectrum. As the total hemoglobin was calculated as the sum of oxyhemoglobin, methemoglobin, and hemichrome, the three main hemoglobin derivatives formed in DBS upon aging, the reflectance spectrum needed to be unmixed to determine the quantity of each of these derivatives. We now simplified the method by only using the reflectance at a single wavelength, located at a quasi-isosbestic point in the reflectance curve. At this wavelength, assuming 1-to-1 stoichiometry of the aging reaction, the reflectance is insensitive to the hemoglobin degradation and only scales with the total amount of hemoglobin and, hence, the Hct. This simplified method was successfully validated. At each quality control level as well as at the limits of quantitation (i.e., 0.20 and 0.67) bias, intra- and interday imprecision were within 10%. Method reproducibility was excellent based on incurred sample reanalysis and surpassed the reproducibility of the original method. Furthermore, the influence of the volume spotted, the measurement location within the spot, as well as storage time and temperature were evaluated, showing no relevant impact of these parameters. Application to 233 patient samples revealed a good correlation between the Hct determined on whole blood and the predicted Hct determined on venous DBS. The bias obtained with Bland and Altman analysis was 0.015 and the limits of agreement were 0.061 and 0.031, indicating that the simplified, noncontact Hct prediction method even outperforms the original method. In addition, using caffeine as a model compound, it was demonstrated that this simplified Hct prediction method can effectively be used to implement a Hct-dependent correction factor to DBS-based results to alleviate the Hct bias.
Original languageEnglish
Pages (from-to)1795-1804
JournalAnalytical Chemistry
Volume90
Issue number3
DOIs
Publication statusPublished - 2018

Cite this