T1 - Delivery of magnetic resonance-guided single-fraction stereotactic lung radiotherapy
AU - Finazzi, Tobias
AU - van Sörnsen de Koste, John R.
AU - Palacios, Miguel A.
AU - Spoelstra, Femke O.B.
AU - Slotman, Berend J.
AU - Haasbeek, Cornelis J.A.
AU - Senan, Suresh
PY - 2020/4/1
Y1 - 2020/4/1
N2 - Background and purpose: Single-fraction stereotactic ablative radiotherapy (SABR) is an effective treatment for early-stage lung cancer, but concerns remain about the accurate delivery of SABR in a single session. We evaluated the delivery of single-fraction lung SABR using magnetic resonance (MR)-guidance. Materials and methods: An MR-simulation was performed in 17 patients, seven of whom were found to be unsuitable, largely due to unreliable tracking of small tumors. Ten patients underwent single-fraction SABR to 34 Gy on a 0.35 T MR-linac system, with online plan adaptation. Gated breath-hold SABR was delivered using a planning target volume (PTV) margin of 5 mm, and a 3 mm gating window. Continuous MR-tracking of the gross tumor volume (GTVt) was performed in sagittal plane, with visual patient feedback provided using an in-room monitor. The real-time MR images were analyzed to determine precision and efficiency of gated delivery. Results: All but one patient completed treatment in a single session. The median total in-room procedure was 120 min, with a median SABR delivery session of 39 min. Review of 7.4 h of cine-MR imaging revealed a mean GTVt coverage by the PTV during beam-on of 99.6%. Breath-hold patterns were variable, resulting in a mean duty cycle efficiency of 51%, but GTVt coverage was not influenced due to real-time MR-guidance. On-table adaptation improved PTV coverage, but had limited impact on GTV doses. Conclusions: Single-fraction gated SABR of lung tumors can be performed with high precision using MR-guidance. However, improvements are needed to ensure MR-tracking of small tumors, and to reduce treatment times.
AB - Background and purpose: Single-fraction stereotactic ablative radiotherapy (SABR) is an effective treatment for early-stage lung cancer, but concerns remain about the accurate delivery of SABR in a single session. We evaluated the delivery of single-fraction lung SABR using magnetic resonance (MR)-guidance. Materials and methods: An MR-simulation was performed in 17 patients, seven of whom were found to be unsuitable, largely due to unreliable tracking of small tumors. Ten patients underwent single-fraction SABR to 34 Gy on a 0.35 T MR-linac system, with online plan adaptation. Gated breath-hold SABR was delivered using a planning target volume (PTV) margin of 5 mm, and a 3 mm gating window. Continuous MR-tracking of the gross tumor volume (GTVt) was performed in sagittal plane, with visual patient feedback provided using an in-room monitor. The real-time MR images were analyzed to determine precision and efficiency of gated delivery. Results: All but one patient completed treatment in a single session. The median total in-room procedure was 120 min, with a median SABR delivery session of 39 min. Review of 7.4 h of cine-MR imaging revealed a mean GTVt coverage by the PTV during beam-on of 99.6%. Breath-hold patterns were variable, resulting in a mean duty cycle efficiency of 51%, but GTVt coverage was not influenced due to real-time MR-guidance. On-table adaptation improved PTV coverage, but had limited impact on GTV doses. Conclusions: Single-fraction gated SABR of lung tumors can be performed with high precision using MR-guidance. However, improvements are needed to ensure MR-tracking of small tumors, and to reduce treatment times.
KW - Adaptive radiotherapy
KW - Lung cancer
KW - MR-guided radiation therapy
KW - SABR
KW - Single-fraction
UR - http://www.scopus.com/inward/record.url?scp=85085054574&partnerID=8YFLogxK
U2 - https://doi.org/10.1016/j.phro.2020.05.002
DO - https://doi.org/10.1016/j.phro.2020.05.002
M3 - Article
C2 - 33458309
SN - 2405-6316
VL - 14
SP - 17
EP - 23
JO - Physics and Imaging in Radiation Oncology
JF - Physics and Imaging in Radiation Oncology
ER -