Abstract

Objective: Cemented total hip arthroplasty (THA) demonstrates superior survival rates compared to uncemented procedures. Nevertheless, most younger patients opt for uncemented THA, as removing well-fixed bone cement in the femur during revisions is complex, particularly the distal cement plug. This removal procedure often increases the risk of femoral fracture or perforation, haemorrhage and weakening bone due to poor drill control and positioning. Aim of this study was to design a novel drill guide to improve drill positioning. Methods and procedures: A novel orthopaedic drill guide was developed, featuring a compliant centralizer activated by a drill guide actuator. Bone models were prepared to assess centralizing performance. Three conditions were tested: drilling without guidance, guided drilling with centralizer activation held, and guided drilling with centralizer activation released. Deviations from the bone centre were measured at the entry and exit point of the drill. Results: In the centralizing performance test, the drill guide significantly reduced drill hole deviations in both entry and exit points compared to the control (p < 0.05). The absolute deviation on the exit side of the cement plug was 10.59mm (SD 1.56) for the 'No drill guide' condition, 3.02mm (SD 2.09) for 'Drill guide - hold' and 2.12mm (SD 1.71) for 'Drill guide - release'. The compliant drill guide centralizer significantly lowered the risk of cortical bone perforation during intramedullary canal drilling in the bone models due to better control of the cement drill position. Clinical and Translational Impact Statement: The drill guide potentially reduces perioperative risks in cemented femoral stem revision. Future research should identify optimal scenarios for its application.

Original languageEnglish
Pages (from-to)340-347
Number of pages8
JournalIEEE Journal of Translational Engineering in Health and Medicine
Volume12
DOIs
Publication statusPublished - 2024

Keywords

  • Drill positioning
  • guiding mechanisms
  • medical device prototyping
  • novel design
  • sustainable design
  • total hip arthroplasty

Cite this