Determination of propofol in low-volume samples by high-performance liquid chromatography with fluorescence detection

C. A. Knibbe, V. S. Koster, V. H. Deneer, R. M. Stuurman, P. F. Kuks, R. Lange

Research output: Contribution to journalArticleAcademicpeer-review

58 Citations (Scopus)

Abstract

In order to determine propofol in rat whole-blood samples of 50 microl, we developed a rapid, simple and reliable method which is characterized by precipitation of blood elements with acetonitrile and submission of the supernatant to HPLC analysis with fluorescence detection. The method described is linear from 0.4 to 40 mg/l and the relative standard deviations in this concentration range are less than 10%. The limit of quantification proved to be 0.4 mg/l. Blood constituents do not interfere with the assay
Original languageEnglish
Pages (from-to)305-310
JournalJournal of chromatography. B, Biomedical sciences and applications
Volume706
Issue number2
DOIs
Publication statusPublished - 1998