TY - JOUR
T1 - Device orientation of a leadless pacemaker and subcutaneous implantable cardioverter-defibrillator in canine and human subjects and the effect on intrabody communication
AU - Quast, Anne-Floor B. E.
AU - Tjong, Fleur V. Y.
AU - Koop, Brendan E.
AU - Wilde, Arthur A. M.
AU - Knops, Reinoud E.
AU - Burke, Martin C.
PY - 2018
Y1 - 2018
N2 - Aims: The development of communicating modular cardiac rhythm management systems relies on effective intrabody communication between a subcutaneous implantable cardioverter-defibrillator (S-ICD) and a leadless pacemaker (LP), using conducted communication. Communication success is affected by the LP and S-ICD orientation. This study is designed to evaluate the orientation of the LP and S-ICD in canine subjects and measure success and threshold of intrabody communication. To gain more human insights, we will explore device orientation in LP and S-ICD patients. Methods and results: Canine subjects implanted with a prototype S-ICD and LP (both Boston Scientific, MA, USA) with anterior-posterior fluoroscopy images were included in this analysis. For comparison, a retrospective analysis of human S-ICD and LP patients was performed. The angle of the long axis of the LP towards the vertical axis of 0°, and distance between the coil and LP were measured. Twenty-three canine subjects were analysed. Median angle of the LP was 29° and median distance of the S-ICD coil to LP was 0.8 cm. All canine subjects had successful communication. The median communicating threshold was 2.5 V. In the human retrospective analysis, 72 LP patients and 100 S-ICD patients were included. The mean angle of the LP was 56° and the median distance between the S-ICD coil and LP was 4.6 cm. Conclusion: Despite the less favourable LP orientation in canine subjects, all communication attempts were successful. In the human subjects, we observed a greater and in theory more favourable LP angle towards the communication vector. These data suggests suitability of human anatomy for conductive intrabody communication.
AB - Aims: The development of communicating modular cardiac rhythm management systems relies on effective intrabody communication between a subcutaneous implantable cardioverter-defibrillator (S-ICD) and a leadless pacemaker (LP), using conducted communication. Communication success is affected by the LP and S-ICD orientation. This study is designed to evaluate the orientation of the LP and S-ICD in canine subjects and measure success and threshold of intrabody communication. To gain more human insights, we will explore device orientation in LP and S-ICD patients. Methods and results: Canine subjects implanted with a prototype S-ICD and LP (both Boston Scientific, MA, USA) with anterior-posterior fluoroscopy images were included in this analysis. For comparison, a retrospective analysis of human S-ICD and LP patients was performed. The angle of the long axis of the LP towards the vertical axis of 0°, and distance between the coil and LP were measured. Twenty-three canine subjects were analysed. Median angle of the LP was 29° and median distance of the S-ICD coil to LP was 0.8 cm. All canine subjects had successful communication. The median communicating threshold was 2.5 V. In the human retrospective analysis, 72 LP patients and 100 S-ICD patients were included. The mean angle of the LP was 56° and the median distance between the S-ICD coil and LP was 4.6 cm. Conclusion: Despite the less favourable LP orientation in canine subjects, all communication attempts were successful. In the human subjects, we observed a greater and in theory more favourable LP angle towards the communication vector. These data suggests suitability of human anatomy for conductive intrabody communication.
UR - https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85053941714&origin=inward
UR - https://www.ncbi.nlm.nih.gov/pubmed/29462311
U2 - https://doi.org/10.1093/europace/euy019
DO - https://doi.org/10.1093/europace/euy019
M3 - Article
C2 - 29462311
SN - 1099-5129
VL - 20
SP - 1866
EP - 1871
JO - Europace : European pacing, arrhythmias, and cardiac electrophysiology
JF - Europace : European pacing, arrhythmias, and cardiac electrophysiology
IS - 11
ER -