Epicardial leads in adult cardiac resynchronization therapy recipients: A study on lead performance, durability, and safety

Maurits S. Buiten, Aafke C. van der Heijden, Robert J. M. Klautz, Martin J. Schalij, Lieselot van Erven

Research output: Contribution to journalArticleAcademicpeer-review

19 Citations (Scopus)

Abstract

Background Transvenous left ventricular (LV) lead placement for cardiac resynchronization therapy-defibrillator (CRT-D) delivery is unsuccessful in 8% to 10% of cases. These patients might benefit from an epicardial lead. However, data on long-term epicardial lead performance are scarce. Furthermore, extracting an epicardial lead requires a rethoracotomy. Objective The purpose of this study was to determine data on almost a decade of experience with epicardial leads and investigate the safety of partially leaving this lead in place after device infection. Methods All adult patients receiving an epicardial lead (Medtronic CapSure Epi, model 4968) for CRT-D in the Leiden University Medical Center were included. Leads were implanted during a standalone procedure or in combination with other cardiothoracic procedures. Electrical lead parameters were assessed at implantation and every 6 months thereafter. In case of device infection the epicardial lead was cut off parasternal, just outside the thoracic cavity, leaving the distal part of the lead in place. Results Two-hundred sixteen patients were included with a median follow-up of 3 years (25th-75th percentile 1.0-5.5). LV pacing threshold decreased within 6 months after implantation [1.1 V (95% confidence interval [CI] 0.9-1.2) vs 0.8 V (95% CI 0.7-0.9), P =.01] and stabilized thereafter. Mean LV electrogram was 15.2 ± 7.5mV, and average lead impedance was 633.5 ± 174.0 Ω. Five-year cumulative incidence was 1.6% for lead failure and 9.6% for device infection. The retained epicardial lead caused skin erosion in 3 patients and fistula formation in 1. Conclusion This study demonstrates that epicardial LV leads have an excellent long-term performance. Partially retaining the lead after device infection was associated with a risk of reinfection with limited long-term clinical implications for the patient.
Original languageEnglish
Pages (from-to)533-539
JournalHeart Rhythm
Volume12
Issue number3
DOIs
Publication statusPublished - 2015
Externally publishedYes

Cite this