T1 - High-Resolution Imaging of Interaction Between Thrombus and Stent-Retriever in Patients With Acute Ischemic Stroke
AU - Autar, Anouchska S A
AU - Hund, Hajo M
AU - Ramlal, Sharad A
AU - Hansen, Daniël
AU - Lycklama À Nijeholt, Geert J
AU - Emmer, Bart J
AU - de Maat, Moniek P M
AU - Dippel, Diederik W J
AU - van der Lugt, Aad
AU - van Es, Adriaan C G M
AU - van Beusekom, Heleen M M
AU - MR CLEAN Registry investigators.
AU - Bot, JCJ
AU - MR CLEAN Registry Investigators
AU - Jansen, Ivo G. H.
AU - Mulder, Maxim J. H. L.
AU - Goldhoorn, Robert-Jan B.
AU - Roozenbeek, Bob
AU - Schonewille, Wouter J.
AU - van den Berg, René
AU - Coutinho, Jonathan M.
AU - Staals, Julie
AU - Postma, Alida A.
AU - Hofmeijer, Jeannette
AU - Martens, Jasper M.
AU - den Hertog, Heleen M.
AU - Sturm, Emiel J. C.
AU - van der Worp, H. Bart
AU - Lo, Rob H.
AU - Heijboer, Roel J. J.
AU - Keizer, Koos
AU - Uyttenboogaart, Maarten
AU - Eshghi, Omid
AU - Wermer, Marieke J. H.
AU - van Walderveen, Marianne A. A.
AU - van Dijk, Ewoud J.
AU - Boogaarts, Hieronymus D.
AU - de Bruijn, Sebastiaan F.
AU - van Dijk, Lukas C.
AU - van den Berg, Jan S. P.
AU - van Hasselt, Boudewijn A. A. M.
AU - de Kort, Paul L. M.
AU - Peluso, Jo P. P.
AU - Schreuder, Tobien H. C. M. L.
AU - Aerden, Leo A. M.
AU - Sprengers, Marieke E. S.
AU - Roosendaal, Stefan D.
AU - Beenen, Ludo F. M.
AU - Roos, Yvo B. W. E. M.
AU - Majoie, Charles B. L. M.
PY - 2018/6/22
Y1 - 2018/6/22
N2 - BACKGROUND: Currently, acute ischemic stroke is still a leading cause of mortality and morbidity. Approximately 2 years ago, mechanical thrombectomy was proven beneficial as a revolutionary new therapy for stroke in the MR-CLEAN trial (A Multicenter Randomized Clinical Trial of Endovascular Treatment for Acute Ischemic Stroke in the Netherlands). However, the mechanisms by which the thrombectomy device, or stent-retriever, interacts with the thrombus are largely unknown. A better understanding could lead to improved efficacy of mechanical thrombectomy devices.METHODS AND RESULTS: Seven stent-retrievers with thrombi still entrapped were collected directly after thrombectomy. The stent-retrievers were studied using micro computed tomography, followed by scanning electron microscopy and light microscopy. Two independent observers rated interaction type and thrombus surface structure (porous filamentous or dense) at the interaction sites.A total of 79 interaction sites between thrombus and stent-retriever were categorized. Thrombus-stent-retriever interaction was found to be adhesive (n=44; 56%) or mechanical (n=35; 44%). Adhesive interaction was most frequently observed at interaction sites with a dense surface, compared with interaction sites with a porous filamentous fibrin surface (38/58; 66% versus 6/21; 29%, P=0.011).CONCLUSIONS: The interaction between thrombus and stent-retriever was predominantly adhesive, not mechanical. Adhesive interaction was strongly associated with the presence of a dense thrombus surface without a porous filamentous fibrin network.
AB - BACKGROUND: Currently, acute ischemic stroke is still a leading cause of mortality and morbidity. Approximately 2 years ago, mechanical thrombectomy was proven beneficial as a revolutionary new therapy for stroke in the MR-CLEAN trial (A Multicenter Randomized Clinical Trial of Endovascular Treatment for Acute Ischemic Stroke in the Netherlands). However, the mechanisms by which the thrombectomy device, or stent-retriever, interacts with the thrombus are largely unknown. A better understanding could lead to improved efficacy of mechanical thrombectomy devices.METHODS AND RESULTS: Seven stent-retrievers with thrombi still entrapped were collected directly after thrombectomy. The stent-retrievers were studied using micro computed tomography, followed by scanning electron microscopy and light microscopy. Two independent observers rated interaction type and thrombus surface structure (porous filamentous or dense) at the interaction sites.A total of 79 interaction sites between thrombus and stent-retriever were categorized. Thrombus-stent-retriever interaction was found to be adhesive (n=44; 56%) or mechanical (n=35; 44%). Adhesive interaction was most frequently observed at interaction sites with a dense surface, compared with interaction sites with a porous filamentous fibrin surface (38/58; 66% versus 6/21; 29%, P=0.011).CONCLUSIONS: The interaction between thrombus and stent-retriever was predominantly adhesive, not mechanical. Adhesive interaction was strongly associated with the presence of a dense thrombus surface without a porous filamentous fibrin network.
UR - https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85049695274&origin=inward
UR - https://www.ncbi.nlm.nih.gov/pubmed/29934420
U2 - https://doi.org/10.1161/JAHA.118.008563
DO - https://doi.org/10.1161/JAHA.118.008563
M3 - Article
C2 - 29934420
SN - 2047-9980
VL - 7
JO - Journal of the American Heart Association
JF - Journal of the American Heart Association
IS - 13
M1 - e008563
ER -