Holmium-166 poly(L-lactic acid) microsphere radioembolisation of the liver: technical aspects studied in a large animal model

M. A. D. Vente, T. C. de Wit, M. A. A. J. van den Bosch, W. Bult, P. R. Seevinck, B. A. Zonnenberg, H. W. A. M. de Jong, G. C. Krijger, C. J. G. Bakker, A. D. van het Schip, J. F. W. Nijsen

Research output: Contribution to journalArticleAcademicpeer-review

37 Citations (Scopus)

Abstract

OBJECTIVE: To assess the accuracy of a scout dose of holmium-166 poly(L-lactic acid) microspheres ((166)Ho-PLLA-MS) in predicting the distribution of a treatment dose of (166)Ho-PLLA-MS, using single photon emission tomography (SPECT). METHODS: A scout dose (60 mg) was injected into the hepatic artery of five pigs and SPECT acquired. Subsequently, a 'treatment dose' was administered (540 mg) and SPECT, computed tomography (CT) and magnetic resonance imaging (MRI) of the total dose performed. The two SPECT images of each animal were compared. To validate quantitative SPECT an ex vivo liver was instilled with (166)Ho-PLLA-MS and SPECT acquired. The liver was cut into slices and planar images were acquired, which were registered to the SPECT image. RESULTS: Qualitatively, the scout dose and total dose images were similar, except in one animal because of catheter displacement. Quantitative analysis, feasible in two animals, tended to confirm this similarity (r(2) = 0.34); in the other animal the relation was significantly better (r(2) = 0.66). The relation between the SPECT and planar images acquired from the ex vivo liver was strong (r(2) = 0.90). CONCLUSION: In the porcine model a scout dose of (166)Ho-PLLA-MS can accurately predict the biodistribution of a treatment dose. Quantitative (166)Ho SPECT was validated for clinical application
Original languageEnglish
Pages (from-to)862-869
JournalEuropean Radiology
Volume20
Issue number4
DOIs
Publication statusPublished - 2010

Cite this