Identification of neisseria gonorrhoeae by the bruker biotyper matrix- assisted laser desorption ionization-time of flight mass spectrometry system is improved by a database extension

Valentijn A. Schweitzer, Alje P. van Dam, I. Putu Yuda Hananta, Rob Schuurman, Johannes G. Kusters, Rob J. Rentenaar

Research output: Contribution to journalArticleAcademicpeer-review

6 Citations (Scopus)

Abstract

Identification of Neisseria gonorrhoeae by the Bruker matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) system may be affected by "B consistency categorization." A supplementary database of 17 N. gonorrhoeae main spectra was constructed. Twelve of 64 N. gonorrhoeae identifications were categorized with B consistency, which disappeared using the supplementary database. Database extension did not result in misidentification of Neisseria meningitidis.
Original languageEnglish
Pages (from-to)1130-1132
JournalJournal of clinical microbiology
Volume54
Issue number4
DOIs
Publication statusPublished - 2016

Cite this