Prosthetic heart valve assessment with multidetector-row CT: imaging characteristics of 91 valves in 83 patients

Jesse Habets, Petr Symersky, Lex A. van Herwerden, Bas A. J. M. de Mol, Anje M. Spijkerboer, Willem P. Th M. Mali, Ricardo P. J. Budde

Research output: Contribution to journalArticleAcademicpeer-review

51 Citations (Scopus)

Abstract

Multidetector CT (MDCT) has shown potential for prosthetic heart valve (PHV) assessment. We assessed the image quality of different PHV types to determine which valves are suitable for MDCT evaluation. All ECG-gated CTs performed in our institutions since 2003 were reviewed for the presence of PHVs. After reconstruction in 3 specific PHV planes, image quality of the supravalvular, perivalvular, subvalvular and valvular regions was scored on a four-point scale (1 = non-diagnostic, 2 = moderate, 3 = good and 4 = excellent) by two independent observers. Eighty-four CT examinations (66 cardiac, 18 limited-dose aortic protocols) of 83 patients with a total of 91 PHVs in the aortic (n = 71), mitral (n = 17), pulmonary (n = 1) and tricuspid (n = 2) position were included. CT was performed on a 16-slice (n = 4), 64-slice (n = 28) or 256-slice (n = 52) MDCT system. Median image quality scores for the supra-, peri- and subvalvular regions and valvular detail were (3.5, 3.3, 3.5 and 3.5, respectively) for bileaflet PHV; (3.0, 3.0, 3.5 and 3.0, respectively) for Medtronic Hall PHV; (1.0, 1.0, 1.0 and 1.0, respectively) for Björk-Shiley and Sorin monoleaflet PHV and (3.5, 3.5, 4.0 and 2.0 respectively) for biological PHV. Currently implanted PHVs have good image quality on MDCT and are suitable for MDCT evaluation
Original languageEnglish
Pages (from-to)1390-1396
JournalEuropean Radiology
Volume21
Issue number7
DOIs
Publication statusPublished - 2011

Cite this