Quantitative comparison of the OCT imaging depth at 1300 nm and 1600 nm

V. M. Kodach, J. Kalkman, D. J. Faber, T. G. van Leeuwen

Research output: Contribution to journalArticleAcademicpeer-review

91 Citations (Scopus)

Abstract

One of the present challenges in optical coherence tomography (OCT) is the visualization of deeper structural morphology in biological tissues. Owing to a reduced scattering, a larger imaging depth can be achieved by using longer wavelengths. In this work, we analyze the OCT imaging depth at wavelengths around 1300 nm and 1600 nm by comparing the scattering coefficient and OCT imaging depth for a range of Intralipid concentrations at constant water content. We observe an enhanced OCT imaging depth for 1600 nm compared to 1300 nm for Intralipid concentrations larger than 4 vol.%. For higher Intralipid concentrations, the imaging depth enhancement reaches 30%. The ratio of scattering coefficients at the two wavelengths is constant over a large range of scattering coefficients and corresponds to a scattering power of 2.8 ± 0.1. Based on our results we expect for biological tissues an increase of the OCT imaging depth at 1600 nm compared to 1300 nm for samples with high scattering power and low water content
Original languageEnglish
Pages (from-to)176-185
JournalBiomedical Optics Express
Volume1
Issue number1
DOIs
Publication statusPublished - 2010

Cite this