Subcutaneous implantable cardioverter-defibrillators: long-term results of the EFFORTLESS study

Pier D. Lambiase, Dominic A. Theuns, Francis Murgatroyd, Craig Barr, Lars Eckardt, Petr Neuzil, Marcoen Scholten, Margaret Hood, J. rgen Kuschyk, Amy J. Brisben, Nathan Carter, Timothy M. Stivland, Reinoud Knops, Lucas V. A. Boersma

Research output: Contribution to journalArticleAcademicpeer-review

45 Citations (Scopus)

Abstract

AIMS: To report 5-year outcomes of EFFORTLESS registry patients with early generation subcutaneous implantable cardioverter-defibrillator (S-ICD) devices. METHODS AND RESULTS: Kaplan-Meier, trend and multivariable analyses were performed for mortality and late (years 2-5) complications, appropriate shock (AS) and inappropriate shock (IAS) rates. Nine hundred and eighty-four of 994 enrolled patients with diverse diagnoses (28% female, 48 ± 17 years, body mass index 27 ± 6 kg/m2, ejection fraction 43 ± 18%) underwent S-ICD implantation. Median follow-up was 5.1 years (interquartile range 4.7-5.5 years). All-cause mortality was 9.3% (95% confidence interval 7.2-11.3%) at 5 years; 703 patients remained in follow-up on study completion, 171 withdrew including 87 (8.8%) with device explanted, and 65 (6.6%) lost to follow-up. Of the explants, only 20 (2.0%) patients needed a transvenous device for pacing indications. First and final shock efficacy for discrete ventricular arrhythmias was consistent at 90% and 98%, respectively, with storm episode final shock efficacy at 95.2%. Time to therapy remained unaltered. Overall 1- and 5-year complication rates were 8.9% and 15.2%, respectively. Early complications did not predict later complications. There were no structural lead failures. Inappropriate shock rates at 1 and 5 years were 8.7% and 16.9%, respectively. Self-terminating inappropriately sensed episodes predicted late IAS. Predictors of late AS included self-terminating appropriately sensed episodes and earlier AS. CONCLUSION: In this diverse S-ICD registry population, spontaneous shock efficacy was consistently high over 5 years. Very few patients underwent S-ICD replacement with a transvenous device for pacing indications. Treated and self-terminating arrhythmic episodes predict future shock events, which should encourage more personalized device optimization.
Original languageEnglish
Pages (from-to)2037-2050
Number of pages14
JournalEuropean Heart journal
Volume43
Issue number21
DOIs
Publication statusPublished - 1 Jun 2022

Keywords

  • Implantable cardioverter-defibrillator
  • Primary prevention
  • Secondary prevention
  • Subcutaneous ICD
  • Sudden death

Cite this