TY - JOUR
T1 - Techniques of Cardiac Output Measurement During Liver Transplantation: Arterial Pulse Wave Versus Thermodilution
AU - Nissen, Peter
AU - van Lieshout, Johannes J.
AU - Novovic, Srdan
AU - Bundgaard-Nielsen, Morten
AU - Secher, Niels H.
PY - 2009
Y1 - 2009
N2 - In this study, we compared continuous cardiac output (CO) obtained from the femoral arterial pressure by simulation of an aortic input impedance model [model-simulated cardiac output (MCO)] to thermodilution cardiac output (TDCO) determined by bolus injection during liver transplantation. Both variables were measured in 39 adult patients (13 females) every 10th minute during liver transplant surgery. Paired measurements were compared during the 4 phases of surgery-dissection, anhepatic phase, early reperfusion (the first 15 minutes after reperfusion), and late reperfusion (15-60 minutes after reperfusion)-without the detection of any significant difference between the 2 estimates of CO. TDCO ranged from 2.3 to 17.2 L/minute, and the bias (the mean difference between MCO and TDCO) prior to calibration was -0.4 +/- 1.6 L/minute (mean +/- standard deviation; 1309 paired measurements; 95% limits of agreement: -3.4 to 2.6 L/minute). After calibration of the first determined MCO by the simultaneously determined TDCO, the bias was 0.1 +/- 1.5 L/minute, with 57% (n = 744) of the comparisons being less than 1 L/minute and 35% (n = 453) being less than 0.5 L/minute; this was independent of the level of CO, and the mutual correlation coefficient was 0.812 (P <0.001). This study indicates that during liver transplantation surgery, MCO reflects TDCO throughout the operation. Thus, for CO, this less invasive method appears to provide a reliable uninterrupted measurement during orthotopic liver transplantation. Liver Transpl 15:287-291, 2009. (C) 2009 AASLD
AB - In this study, we compared continuous cardiac output (CO) obtained from the femoral arterial pressure by simulation of an aortic input impedance model [model-simulated cardiac output (MCO)] to thermodilution cardiac output (TDCO) determined by bolus injection during liver transplantation. Both variables were measured in 39 adult patients (13 females) every 10th minute during liver transplant surgery. Paired measurements were compared during the 4 phases of surgery-dissection, anhepatic phase, early reperfusion (the first 15 minutes after reperfusion), and late reperfusion (15-60 minutes after reperfusion)-without the detection of any significant difference between the 2 estimates of CO. TDCO ranged from 2.3 to 17.2 L/minute, and the bias (the mean difference between MCO and TDCO) prior to calibration was -0.4 +/- 1.6 L/minute (mean +/- standard deviation; 1309 paired measurements; 95% limits of agreement: -3.4 to 2.6 L/minute). After calibration of the first determined MCO by the simultaneously determined TDCO, the bias was 0.1 +/- 1.5 L/minute, with 57% (n = 744) of the comparisons being less than 1 L/minute and 35% (n = 453) being less than 0.5 L/minute; this was independent of the level of CO, and the mutual correlation coefficient was 0.812 (P <0.001). This study indicates that during liver transplantation surgery, MCO reflects TDCO throughout the operation. Thus, for CO, this less invasive method appears to provide a reliable uninterrupted measurement during orthotopic liver transplantation. Liver Transpl 15:287-291, 2009. (C) 2009 AASLD
U2 - https://doi.org/10.1002/lt.21689
DO - https://doi.org/10.1002/lt.21689
M3 - Article
C2 - 19242994
SN - 1527-6465
VL - 15
SP - 287
EP - 291
JO - Liver transplantation
JF - Liver transplantation
IS - 3
ER -