TY - JOUR
T1 - The learning curve associated with the implantation of the Nanostim leadless pacemaker
AU - on behalf of the Leadless II IDE and
AU - Tjong, Fleur V. Y.
AU - Beurskens, Niek E. G.
AU - Neuzil, Petr
AU - Defaye, Pascal
AU - Delnoy, Peter-Paul
AU - Ip, John
AU - Guerrero, Juan Jose Garcia
AU - Rashtian, Mayer
AU - Banker, Rajesh
AU - Reddy, Vivek
AU - Exner, Derek
AU - Sperzel, Johannes
AU - Knops, Reinoud E.
PY - 2018
Y1 - 2018
N2 - Purpose: Use of novel medical technologies, such as leadless pacemaker (LP) therapy, may be subjected to a learning curve effect. The objective of the current study was to assess the impact of operators’ experience on the occurrence of serious adverse device effects (SADE) and procedural efficiency. Methods: Patients implanted with a Nanostim LP (Abbott, USA) within two prospective studies (i.e., LEADLESS ll IDE and Leadless Observational Study) were assessed. Patients were categorized into quartiles based on operator experience. Learning curve analysis included the comparison of SADE rates at 30 days post-implant per quartile and between patients in quartile 4 (> 10 implants) and patients in quartiles 1 through 3 (1–10 implants). Procedural efficiency was assessed based on procedure duration and repositioning attempts. Results: Nanostim LP implant was performed in 1439 patients by 171 implanters at 60 centers in 10 countries. A total of 91 (6.4%) patients experienced a SADE in the first 30 days. SADE rates dropped from 7.4 to 4.5% (p = 0.038) after more than 10 implants per operator. Total procedure duration decreased from 30.9 ± 19.1 min in quartile 1 to 21.6 ± 13.2 min (p < 0.001) in quartile 4. The need for multiple repositionings during the LP procedure reduced in quartile 4 (14.8%), compared to quartiles 1 (26.8%; p < 0.001), 2 (26.6%; p < 0.001), and 3 (20.4%; p = 0.03). Conclusions: Learning curves exist for Nanostim LP implantation. Procedure efficiency improved with increased operator experience, according to a decrease in the incidence of SADE, procedure duration, and repositioning attempts.
AB - Purpose: Use of novel medical technologies, such as leadless pacemaker (LP) therapy, may be subjected to a learning curve effect. The objective of the current study was to assess the impact of operators’ experience on the occurrence of serious adverse device effects (SADE) and procedural efficiency. Methods: Patients implanted with a Nanostim LP (Abbott, USA) within two prospective studies (i.e., LEADLESS ll IDE and Leadless Observational Study) were assessed. Patients were categorized into quartiles based on operator experience. Learning curve analysis included the comparison of SADE rates at 30 days post-implant per quartile and between patients in quartile 4 (> 10 implants) and patients in quartiles 1 through 3 (1–10 implants). Procedural efficiency was assessed based on procedure duration and repositioning attempts. Results: Nanostim LP implant was performed in 1439 patients by 171 implanters at 60 centers in 10 countries. A total of 91 (6.4%) patients experienced a SADE in the first 30 days. SADE rates dropped from 7.4 to 4.5% (p = 0.038) after more than 10 implants per operator. Total procedure duration decreased from 30.9 ± 19.1 min in quartile 1 to 21.6 ± 13.2 min (p < 0.001) in quartile 4. The need for multiple repositionings during the LP procedure reduced in quartile 4 (14.8%), compared to quartiles 1 (26.8%; p < 0.001), 2 (26.6%; p < 0.001), and 3 (20.4%; p = 0.03). Conclusions: Learning curves exist for Nanostim LP implantation. Procedure efficiency improved with increased operator experience, according to a decrease in the incidence of SADE, procedure duration, and repositioning attempts.
UR - https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85052127123&origin=inward
UR - https://www.ncbi.nlm.nih.gov/pubmed/30105428
U2 - https://doi.org/10.1007/s10840-018-0438-8
DO - https://doi.org/10.1007/s10840-018-0438-8
M3 - Article
C2 - 30105428
SN - 1383-875X
VL - 53
SP - 239
EP - 247
JO - Journal of Interventional Cardiac Electrophysiology
JF - Journal of Interventional Cardiac Electrophysiology
IS - 2