Troubleshooting in Transatrial Tricuspid Valve-in-Valve Implantation

Research output: Contribution to journalEditorialAcademicpeer-review

12 Citations (Scopus)

Abstract

The established treatment for degenerated stenotic tricuspid bioprostheses is reoperation. Recently, transcatheter tricuspid valve implantation has been reported as an alternative option. This case report describes a complex transcatheter tricuspid valve implantation in a degenerated Medtronic intact 31 mm bioprosthesis. Implantation of a 26 mm Edwards Sapien valve failed, subsequent transcatheter implantation of a 29 mm Edwards Sapien valve was successful. (Ann Thorac Surg 2012;94:1349-52) (C) 2012 by The Society of Thoracic Surgeons
Original languageEnglish
Pages (from-to)1349-1352
JournalAnnals of Thoracic Surgery
Volume94
Issue number4
DOIs
Publication statusPublished - 2012