T1 - Updating PET/CT performance standards and PET/CT interpretation criteria should go hand in hand
AU - Boellaard, Ronald
AU - Sera, Terez
AU - Kaalep, Andres
AU - Hoekstra, Otto S.
AU - Barrington, Sally F.
AU - Zijlstra, Josée M.
PY - 2019/1/1
Y1 - 2019/1/1
N2 - This letter aims at explaining that adjusting the performance of PET/CT systems to a new standard also requires updating of interpretation criteria. Simply changing one aspect of the imaging procedure, i.e., PET/CT performance and image quality, and not adapting interpretation criteria will result in an increase of false positive (or negative) reads.
AB - This letter aims at explaining that adjusting the performance of PET/CT systems to a new standard also requires updating of interpretation criteria. Simply changing one aspect of the imaging procedure, i.e., PET/CT performance and image quality, and not adapting interpretation criteria will result in an increase of false positive (or negative) reads.