@article{4870476c5459400b80102b5bff5a3869,
title = "Ventilating two patients with one ventilator: technical setup and laboratory testing",
abstract = "With a modified circuit, it is feasible to ventilate two patients with one ventilator over a relevant range of compliances. Adding inspiratory resistance allows individual titration of tidal volume, and incorporating one-way valves prevents pendelluft. https://bit.ly/3ex8SYP.",
author = "{de Jongh}, {Frans H C} and {de Vries}, {Heder J} and Warnaar, {Rob S P} and Eline Oppersma and Rudolf Verdaasdonk and Heunks, {Leo M A} and Jonne Doorduin",
note = "Copyright {\textcopyright}ERS 2020.",
year = "2020",
month = apr,
doi = "https://doi.org/10.1183/23120541.00256-2020",
language = "English",
volume = "6",
journal = "BMJ Open",
issn = "2312-0541",
publisher = "BMJ Publishing Group",
number = "2",
}