Closed–loop oxygen control improves oxygenation in pediatric patients under high–flow nasal oxygen—A randomized crossover study

Ozlem Sandal, Gokhan Ceylan, Sevgi Topal, Pinar Hepduman, Mustafa Colak, Dominik Novotni, Ekin Soydan, Utku Karaarslan, Gulhan Atakul, Marcus J. Schultz, Hasan Ağın

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

Background: We assessed the effect of a closed–loop oxygen control system in pediatric patients receiving high–flow nasal oxygen therapy (HFNO). Methods: A multicentre, single–blinded, randomized, and cross–over study. Patients aged between 1 month and 18 years of age receiving HFNO for acute hypoxemic respiratory failure (AHRF) were randomly assigned to start with a 2–h period of closed–loop oxygen control or a 2–h period of manual oxygen titrations, after which the patient switched to the alternative therapy. The endpoints were the percentage of time spent in predefined SpO2 ranges (primary), FiO2, SpO2/FiO2, and the number of manual adjustments. Findings: We included 23 patients, aged a median of 18 (3–26) months. Patients spent more time in a predefined optimal SpO2 range when the closed–loop oxygen controller was activated compared to manual oxygen titrations [91⋅3% (IQR 78⋅4–95⋅1%) vs. 63⋅0% (IQR 44⋅4–70⋅7%)], mean difference [28⋅2% (95%–CI 20⋅6–37⋅8%); P < 0.001]. Median FiO2 was lower [33⋅3% (IQR 26⋅6–44⋅6%) vs. 42⋅6% (IQR 33⋅6–49⋅9%); P = 0.07], but median SpO2/FiO2 was higher [289 (IQR 207–348) vs. 194 (IQR 98–317); P = 0.023] with closed–loop oxygen control. The median number of manual adjustments was lower with closed–loop oxygen control [0⋅0 (IQR 0⋅0–0⋅0) vs. 0⋅5 (IQR 0⋅0–1⋅0); P < 0.001]. Conclusion: Closed-loop oxygen control improves oxygenation therapy in pediatric patients receiving HFNO for AHRF and potentially leads to more efficient oxygen use. It reduces the number of manual adjustments, which may translate into decreased workloads of healthcare providers. Clinical trial registration: [www.ClinicalTrials.gov], identifier [NCT 05032365].

Original languageEnglish
Article number1046902
JournalFrontiers in Medicine
Volume9
DOIs
Publication statusPublished - 16 Nov 2022

Keywords

  • automation
  • closed-loop
  • high flow (NHF)
  • hypoxemia
  • intensive care
  • oxygen controller
  • oxygen therapy
  • pediatric [MeSH]

Cite this