Abstract

Accurate transfer of a preoperatively planned osteotomy plane to the bone is of significance for corrective surgery, tumor resection, implant positioning and evaluation of new osteotomy techniques. Methods for comparing a preoperatively planned osteotomy plane with a surgical cut exist but the accuracy of these techniques are either limited or unknown. This paper proposes and evaluates a CT-based technique that enables comparing virtual with actual osteotomy planes. The methodological accuracy and reproducibility of the technique is evaluated using CT-derived volume data of a cadaver limb, which serves to plan TKA osteotomies in 3-D space and to simulate perfect osteotomies not hampered by surgical errors. The methodological variability of the technique is further investigated with repeated CT scans after actual osteotomy surgery of the same cadaver specimen. Plane displacement (derr) and angulation errors in the sagittal and coronal plane (βerr, γerr) are measured with high accuracy and reproducibility (derr=-0.11±0.06mm; βerr=0.08±0.04°, γerr=-0.03±0.03°). The proposed method for evaluating an osteotomy plane position and orientation has a high intrinsic accuracy and reproducibility. The method can be of great value for measuring the transfer accuracy of new techniques for positioning and orienting a surgical cut in 3-D space
Original languageEnglish
Pages (from-to)1081-1087
JournalMedical engineering & physics
Volume36
Issue number8
DOIs
Publication statusPublished - 2014