TY - JOUR
T1 - Mitigating SUV uncertainties using total body PET imaging
AU - Smith, Charlotte L. C.
AU - Zwezerijnen, Gerben J. C.
AU - den Hollander, Marijke E.
AU - Weijland, Jolijn
AU - Yaqub, Maqsood
AU - Boellaard, Ronald
N1 - Funding Information: RB received a research grant from Siemens Healthineers. No other potential conflicts of interest relevant to this article exist. Publisher Copyright: © 2023, The Author(s).
PY - 2023
Y1 - 2023
N2 - Purpose: Standardised uptake values (SUV) are commonly used to quantify 18F-FDG lesion uptake. However, SUVs may suffer from several uncertainties and errors. Long-axial field-of-view (LAFOV) PET/CT systems might enable image-based quality control (QC) by deriving 18F-FDG activity and weight from total body (TB) 18F-FDG PET images. In this study, we aimed to develop these image-based QC to reduce errors and mitigate SUV uncertainties. Methods: Twenty-five out of 81 patient scans from a LAFOV PET/CT system were used to determine regression fits for deriving of image-derived activity and weight. Thereafter, the regression fits were applied to 56 independent 18F-FDG PET scans from the same scanner to determine if injected activity and weight could be obtained accurately from TB and half-body (HB) scans. Additionally, we studied the impact of image-based values on the precision of liver SUVmean and lesion SUVpeak. Finally, 20 scans were acquired from a short-axial field-of-view (SAFOV) PET/CT system to determine if the regression fits also applied to HB scans from a SAFOV system. Results: Both TB and HB 18F-FDG activity and weight significantly predicted reported injected activity (r = 0.999; r = 0.984) and weight (r = 0.999; r = 0.987), respectively. After applying the regression fits, 18F-FDG activity and weight were accurately derived within 4.8% and 3.2% from TB scans and within 4.9% and 3.1% from HB, respectively. Image-derived values also mitigated liver and lesion SUV variability compared with reported values. Moreover, 18F-FDG activity and weight obtained from a SAFOV scanner were derived within 6.7% and 4.5%, respectively. Conclusion: 18F-FDG activity and weight can be derived accurately from TB and HB scans, and image-derived values improved SUV precision and corrected for lesion SUV errors. Therefore, image-derived values should be included as QC to generate a more reliable and reproducible quantitative uptake measurement.
AB - Purpose: Standardised uptake values (SUV) are commonly used to quantify 18F-FDG lesion uptake. However, SUVs may suffer from several uncertainties and errors. Long-axial field-of-view (LAFOV) PET/CT systems might enable image-based quality control (QC) by deriving 18F-FDG activity and weight from total body (TB) 18F-FDG PET images. In this study, we aimed to develop these image-based QC to reduce errors and mitigate SUV uncertainties. Methods: Twenty-five out of 81 patient scans from a LAFOV PET/CT system were used to determine regression fits for deriving of image-derived activity and weight. Thereafter, the regression fits were applied to 56 independent 18F-FDG PET scans from the same scanner to determine if injected activity and weight could be obtained accurately from TB and half-body (HB) scans. Additionally, we studied the impact of image-based values on the precision of liver SUVmean and lesion SUVpeak. Finally, 20 scans were acquired from a short-axial field-of-view (SAFOV) PET/CT system to determine if the regression fits also applied to HB scans from a SAFOV system. Results: Both TB and HB 18F-FDG activity and weight significantly predicted reported injected activity (r = 0.999; r = 0.984) and weight (r = 0.999; r = 0.987), respectively. After applying the regression fits, 18F-FDG activity and weight were accurately derived within 4.8% and 3.2% from TB scans and within 4.9% and 3.1% from HB, respectively. Image-derived values also mitigated liver and lesion SUV variability compared with reported values. Moreover, 18F-FDG activity and weight obtained from a SAFOV scanner were derived within 6.7% and 4.5%, respectively. Conclusion: 18F-FDG activity and weight can be derived accurately from TB and HB scans, and image-derived values improved SUV precision and corrected for lesion SUV errors. Therefore, image-derived values should be included as QC to generate a more reliable and reproducible quantitative uptake measurement.
KW - Body weight
KW - F-FDG PET/CT
KW - Image based
KW - LAFOV PET
KW - Quality control
UR - http://www.scopus.com/inward/record.url?scp=85176268234&partnerID=8YFLogxK
U2 - https://doi.org/10.1007/s00259-023-06503-x
DO - https://doi.org/10.1007/s00259-023-06503-x
M3 - Article
C2 - 37953391
SN - 1619-7070
JO - European journal of nuclear medicine and molecular imaging
JF - European journal of nuclear medicine and molecular imaging
ER -