Purpose: Spine stereotactic body radiation therapy (SBRT) requires high positioning accuracy and a stable patient to maximize target coverage and reduce excessive irradiation to organs at risk. Positional verification during spine SBRT delivery helps to ensure accurate positioning for all patients. We report our experience with noninvasive 3-dimensional target position monitoring during volumetric modulated arc therapy of spine metastases in nonimmobilized patients positioned using only a thin mattress and simple arm and knee supports.
Methods and materials: Fluoroscopic planar kV images were acquired at 7 frames/s using the on-board imaging system during volumetric modulated arc therapy spine SBRT. Template matching and triangulation were used to track the target in vertical, longitudinal, and lateral directions. If the tracking trace deviated >1 mm from the planned position in ≥1 direction, treatment was manually interrupted and 6-dimensional cone beam computed tomography (CBCT)-based couch correction was performed. Tracking data were used to retrospectively analyze the target position. Positional data, agreement with CBCT, correlation between position of the couch and direction of any positional correction, and treatment times were analyzed.
Results: In total, 175 fractions were analyzed. Delivery was interrupted 83 times in 66 fractions for a deviation >1 mm. In 97% of cases the difference between tracking data and subsequent clinical shift performed after the CBCT match was ≤0.5 mm. Lateral/longitudinal shift performed after intervention correlated with the couch roll/pitch at the start of treatment (correlation coefficient, -0.63/0.53). Mean (SD; range) time between start of first imaging and end of the last arc was 15.2 minutes (5.1; 7.6-36.3).
Conclusions: Spine tracking during irradiation can be used to prompt an intervention CBCT scan and repositioning so that a spine SBRT target deviates by ≤1 mm from the planned position, even in nonimmobilized patients. kV tracking and CBCT are in good agreement. The data support verification CBCT after all 6 degrees-of-freedom positional corrections in nonimmobilized spine SBRT patients.
Purpose: Spine stereotactic body radiation therapy (SBRT) requires high positioning accuracy and a stable patient to maximize target coverage and reduce excessive irradiation to organs at risk. Positional verification during spine SBRT delivery helps to ensure accurate positioning for all patients. We report our experience with noninvasive 3-dimensional target position monitoring during volumetric modulated arc therapy of spine metastases in nonimmobilized patients positioned using only a thin mattress and simple arm and knee supports.
Methods and materials: Fluoroscopic planar kV images were acquired at 7 frames/s using the on-board imaging system during volumetric modulated arc therapy spine SBRT. Template matching and triangulation were used to track the target in vertical, longitudinal, and lateral directions. If the tracking trace deviated >1 mm from the planned position in ≥1 direction, treatment was manually interrupted and 6-dimensional cone beam computed tomography (CBCT)-based couch correction was performed. Tracking data were used to retrospectively analyze the target position. Positional data, agreement with CBCT, correlation between position of the couch and direction of any positional correction, and treatment times were analyzed.
Results: In total, 175 fractions were analyzed. Delivery was interrupted 83 times in 66 fractions for a deviation >1 mm. In 97% of cases the difference between tracking data and subsequent clinical shift performed after the CBCT match was ≤0.5 mm. Lateral/longitudinal shift performed after intervention correlated with the couch roll/pitch at the start of treatment (correlation coefficient, -0.63/0.53). Mean (SD; range) time between start of first imaging and end of the last arc was 15.2 minutes (5.1; 7.6-36.3).
Conclusions: Spine tracking during irradiation can be used to prompt an intervention CBCT scan and repositioning so that a spine SBRT target deviates by ≤1 mm from the planned position, even in nonimmobilized patients. kV tracking and CBCT are in good agreement. The data support verification CBCT after all 6 degrees-of-freedom positional corrections in nonimmobilized spine SBRT patients.