Tailoring four-dimensional cone-beam CT acquisition settings for fiducial marker-based image guidance in radiation therapy

Peng Jin, Niek van Wieringen, Maarten C. C. M. Hulshof, Arjan Bel, Tanja Alderliesten

Research output: Contribution to journalArticleAcademicpeer-review

1 Citation (Scopus)

Abstract

Use of four-dimensional cone-beam CT (4D-CBCT) and fiducial markers for image guidance during radiation therapy (RT) of mobile tumors is challenging due to the trade-off among image quality, imaging dose, and scanning time. This study aimed to investigate different 4D-CBCT acquisition settings for good visibility of fiducial markers in 4D-CBCT. Using these 4D-CBCTs, the feasibility of marker-based 4D registration for RT setup verification and manual respiration-induced motion quantification was investigated. For this, we applied a dynamic phantom with three different breathing motion amplitudes and included two patients with implanted markers. Irrespective of the motion amplitude, for a medium field of view (FOV), marker visibility was improved by reducing the imaging dose per projection and increasing the number of projection images; however, the scanning time was 4 to 8 min. For a small FOV, the total imaging dose and the scanning time were reduced (62.5% of the dose using a medium FOV, 2.5 min) without losing marker visibility. However, the body contour could be missing for a small FOV, which is not preferred in RT. The marker-based 4D setup verification was feasible for both the phantom and patient data. Moreover, manual marker motion quantification can achieve a high accuracy with a mean error of [Formula: see text]
Original languageEnglish
Pages (from-to)021207
JournalJournal of medical imaging (Bellingham, Wash.)
Volume5
Issue number2
DOIs
Publication statusPublished - 2018

Cite this